Some samples start degrading the moment they're exposed to lab light. If your peak areas drop across a sequence, or unexpected degradant peaks appear in later injections, light could be the cause.
The fix starts with the vial. Here's how to choose the right one.
Plenty of common analytes break down under UV and visible light, including:
If you're not sure, check the method, the monograph or the analyte's safety data sheet. When in doubt, run a quick stability test (more on that below).
Amber borosilicate vials block most UV and short-wavelength visible light, which is the range that does the damage. For light-sensitive HPLC samples, they should be your starting point.
What to look for:
Amber glass gets its colour from metal oxides such as iron and titanium. That means slightly higher free ion content than clear 33-expansion glass.
For most work this doesn't matter. But if your analytes are chelating compounds, trace metals, or sensitive to metal adsorption, consider:
The vial is only one stage. Light exposure adds up across the whole process:
Not sure whether your analyte needs protecting? Prepare the same standard in clear and amber vials, leave them in the autosampler, and inject at intervals over your typical run time. If the clear vial's peak area drops or degradant peaks appear, you have your answer and documented evidence for your method file.
Light protection doesn't replace a good seal. Pair amber vials with the right closure for your application. PTFE/silicone septa suit most HPLC work, and Ultra Low Bleed Septa are the choice for LC-MS.
For light-sensitive HPLC samples, amber Type I borosilicate vials are the safe default. Use silanised amber where metal interaction is a risk, and protect the whole workflow, not just the vial.